Understanding basic spectrum analyzer operation

06, May. 2024

 

Understanding basic spectrum analyzer operation

R&S®Essentials | Spectrum analyzers fundamentals

For more information, please visit Vbw And Rbw.

Understanding basic spectrum analyzer operation

Author: Paul Denisowski, Test & measurement expert

The following is an introduction to basic spectrum analyzer operation.

Spectrum analyzers are frequency-domain instruments, showing power versus frequency. This is also the most fundamental measurement on a spectrum analyzer: a plot of power versus frequency.

Most spectrum analyzers automate certain power versus frequency type measurements, like AM modulation depth or third order intercept. These measurements could be done manually but automating them increases efficiency and accuracy. Other measurements like occupied bandwidth or adjacent channel leakage ratio, would be difficult or impossible to manually measure.

There are four essential parameters needed to operate a spectrum analyzer. These four parameters are

  • Center and span
  • Reference level
  • Resolution bandwidth
  • Video bandwidth

These settings are used when making almost any kind of spectrum measurements.

Center and span

Center and span define the frequency range to be measured by setting the stop and start frequencies.

As an example, to measure power between 840 MHz and 860 MHz. These values could be entered into a spectrum analyzer as start and stop frequencies, but center and span are used more commonly. The names are self-explanatory: center is the frequency in the middle of the display, and span is the width of the display. The range 840 MHz to 860 MHz is the same as a center of 850 MHz and a span of 20 MHz. Most often the center frequency of the signal of interest is known, and by using span it’s easier to zoom in and zoom out by just increasing or decreasing the span.

Spectrum Analysis Fundamentals, Part 4: Video Filtering ...

Spectrum Analysis Fundamentals, Part 4: Video Filtering and Trace Averaging

Welcome back to the Spectrum Analysis Basics blog series. In part 3, I went over detector types and how spectrum analyzers gather and display data, including average detection. In this blog, I’ll go over averaging processes. Averaging processes help to smooth the variations in envelope-detected amplitudes. This blog will discuss two of these processes — video filtering and trace averaging.

Video Filtering

Spectrum analyzers display signals as well as their own internal noise. Figure 1 shows an example of a signal plus internal noise.

Figure 1: A spectrum with no filtering, displaying a signal and noise

Smoothing, or averaging, the display helps to reduce the effect of noise on signal amplitude, such as that shown in Figure 2. A variable video filter performs this smoothing.

Figure 2: The display from Figure 1 with filtering

In a spectrum analyzer schematic, the low-pass video filter comes after the envelope detector and determines the bandwidth of the video signal. When a signal passes through the intermediate frequency (IF) chain, the envelope can rapidly vary. To reduce these variations, lower the cutoff frequency of the video filter to a point where it is smaller than resolution bandwidth (RBW). Figure 3 shows the effect of reducing the video bandwidth (VBW) to resolution bandwidth (RBW) ratio.

Figure 3: The effect of a VBW-to-RBW ratio of 3:1 (blue), 1:10 (yellow), and 1:100 (green)

When measuring noise, the effect is most noticeable when using a wide RBW. Peak-to-peak variations of noise minimize as you reduce the VBW. To ensure a smooth signal, stick to ratios of 0.01 or less. Don’t worry – the video filter will not affect parts of the trace that are already smooth, such as a sinusoid that is well above the noise.

The effect of video filtering changes based on your chosen detection mode. In positive peak detection mode, if VBW > RBW, changing the VBW does not make a significant difference in peak-to-peak noise levels. Additionally, if VBW < RBW, changing VBW causes small fluctuations in the noise level, altering the peak values of the smoothed noise envelope. When using average detection, the average noise level remains constant.

Are you interested in learning more about synthesized signal generator for sale? Contact us today to secure an expert consultation!

Trace averaging

In trace averaging, smoothing takes place over two or more sweeps on a point-by-point basis. At each new display point, the spectrum analyzer averages the new value with the previously averaged data using the following formula:

Where

Aavg = new average value

Aprior avg = average from prior sweep

An = measured value on current sweep

n = number of current sweep

To change the degree of smoothing, you can select the number of sweeps over which the averaging occurs. Figure 4 shows the effect of increasing the sweep number, n. While decreasing the VBW affects sweep time, trace averaging has no effect. However, it still takes about the same amount of time as video filtering due to the number of sweeps required.

Figure 4: Trace averaging for 1, 5, 20, and 100 sweeps, from top to bottom

Generally, it does not matter which type of filtering you use. — you will likely reach the same results. However, be sure to note the stark difference between the two: video filtering performs averaging in real time, while trace averaging does not. With video filtering, you see the full effect of the smoothing at each point on the display as the sweep progresses. The analyzer averages each point only once at a rate of about 1/VBW per sweep. Alternatively, with trace averaging, multiple sweeps occur to achieve the full degree of averaging. The averaging at each point takes place of the full time period needed to complete multiple sweeps.

On certain signals, you can get significantly different results depending on your chosen method of filtering. When analyzing a signal with a spectrum that changes with time, your average may be different on each sweep when using video filtering, depicted in Figure 5. If you use trace averaging over many sweeps, the value ends up much closer to the true average, shown in Figure 6.

Figure 5: Video filtering on a signal with a spectrum that changes with time

Figure 6: Trace averaging on the same signal as shown in Figure 5

Conclusion

Averaging helps reduce variations due to noise in your signal, enabling you to differentiate important spectral components from noise. While there are a number of methods out there, make sure to choose the right method for your specific signal when differences in averaging occur. For more background information on spectrum analysis basics, start at Part 1.

Next time, I will go over time gating — an essential topic for spectrum analysis both in classic spectrum analyzers and modern, advanced signal analyzers. For a cumulative overview and in-depth reading on spectrum analysis, check out Application Note 150.

Learn more about Keysight RF spectrum analysis