To begin the Pressure Decay leak test, the test part [TP] is attached to the test port. The tester then pressurizes the part with pressure [+P] to the desired test pressure set by the pressure regulator [R1] during the Fill step by opening valves [V1] and [V2].
Click here to get more.
Upon completion of the Fill step, pressure is trapped inside the part as [V2] closes. After a slight delay, [V1] closes allowing the line between [V1] and [V2] to vent to atmosphere through [V1] allowing any leakage through [V2] to cause the tester to see this leakage and fail all tests. The extra valve allows for a fail-safe configuration.
Trapped pressure inside the test circuit is held through the Stabilize step and then measured by the testers pressure sensor [PS] during the Test step.
If the part exceeds the programmed leak tolerance (pressure drop over time or quantified leak rate), the testers display indicates the reason for the test failure and displays the pressure loss or leak rate in user defined units of measure.
Upon completion of the Test step, pressure trapped in the part is vented to atmosphere through [V1] by opening [V2] during the Vent step, and the tester is ready for the next test.
(3 minute read)
A simpler integrity test for less critical applications is a Pressure Decay Test. This requires the same air (or nitrogen) source and an accurate pressure gauge but does not require a mass flow meter. This type of test can be performed before and after use and is sensitive enough to detect very small leaks.
All test procedures outlined below should be followed when performing pressure decay testing of cartridge filters. The schematic below represents a cartridge integrity test setup with all of the piping, valves, gauges, and regulators needed to properly integrity test filters in place.
NOTE: This blog and the schematics in it show a filter housing with a single 10 cartridge installed. Different-length cartridges can be accommodated using the appropriately sized housing that will also match the end configuration on the cartridge. Contact CPF Customer Service for additional details on multi-filter housing installation.
integrity test parameters
Pore Size (µm) Bubble point (psi) Pressure Decay1 / Diffusion2 Test Pressure (psi) 0.03 * 60 0.10 * 48 0.22 50 35 0.45 25 20 0.65 19 15 0.80 15 12* Bubble Point exceeds maximum operating parameters
1Pressure Decay Test Formula
DP = D * T * Pa / V
Where:
Cell Instruments are exported all over the world and different industries with quality first. Our belief is to provide our customers with more and better high value-added products. Let's create a better future together.
DP = Maximum allowable Pressure Drop; (starting pressure final pressure at end of test interval)
D = Diffusion Rate limit; (from the cartridge data sheet)
T = Time (minutes); typically, 10 minutes
Pa = Atmospheric Pressure (14.7 psi)
V = Upstream Volume of test apparatus (cc); Can be calculated by filling the upstream side with water, then draining through V2 and measuring water volume
2Diffusion specification values are shown on the Data Sheet for the cartridge being tested
Important Notes
ALL connections must be airtight. Any leaks in the plumbing can cause inaccurate readings. It is recommended to leak-check the system on a regular basis.
Marginal integrity failures can often be the result of inadequate cartridge wetting. As indicated in the instructions below, re-wet the module and retest.
Gross failures could be the result of an improperly installed cartridge in the housing or a plumbing leak. Make sure the cartridge is properly installed with any O-Rings undamaged and properly seated. Check the system for air leaks.
Cartridge Installation & Wetting
Pressure decay procedure
If the cartridge appears to fail the initial test, repeat the Cartridge Wetting procedure and increase the rinse time to twice the initial amount to assure that the media has been properly wetted. Then, retest. If necessary, check the cartridge seal in the housing.
Critical Process Filtration, Inc., an ISO -certified manufacturer of process filters has been helping customers for over 20 years to build and improve process filtration systems. Proven products, technical services and support, fast shipping, and very competitive pricing assure companies of getting the best value and expertise. Our comprehensive testing & analysis and validation services support your team whenever they need it. Partnering with your process team is how we deliver your companys solution right the first time. For more information on Integrity Testing in your process, contact us here.
If you are looking for more details, kindly visit pressure decay method.
more details learn more View Details visit our website our website more information Multimeter Frequency Counter What Is a Seismic Sensor What Is Data Harvesting? How Does A Signal Generator Work What Is Seismic Sensor Rbw And Vbw Frequency Counter Application optical components How to Measure Phase Noise by Spectrum Analyzer How to Measure Phase Noise by Spectrum Analyzer